Maintenance of MEG system

Checks to be made

Helium level

  • Every two days check Helium Level it should be higher than 50%

  • Check the ATP and ATL gas flow and pressure
    • If low Helium pressure (Low G FLOW on the ATL) then
      • Remove the hose between ATP and ATL

      • Check if helium is passing through the hose

      • Plug the hose again, see if G Flow increases

      • Restart the ATL from the green button on the rear panel

Data retrieval from ATP and ATL for diagnostic

When the ATP and ATL helium recovery system exhibits low gas flow or unusual temperature/pressure, contact the references below and provide them the following data

  1. Retrieve data from ATP and ATL by opening an FTP connection to 10.224.44.200 and 10.224.44.201

  2. Connect to the NYU Abu Dhabi VPN

  3. Use the following information username: qd, password: 79653

  4. Navigate to the /StorageCard/DAT_Files/ directory for each of the previous IP addresses, send the latest .dat files

Note

documentation in the link: QD Documentation for Data retrieval Sheet to be filled by maintenance team for Helium levels Helium Filling Sheet

Check air pressure every month

  • The air pressure inside the meg can be not so good, it must be 0.7, 0.8 this could be a compressor problem

  • When the air pressure is low, the door could be stuck

MSR Door:

  • Test the emergency button every week

  • Test if the pressure release when using the emergency button is getting heavier or not releasing pressure as it is supposed to be

  • If using the manual handle, make sure that to reset the door, you need to put the handle in the original position or else the door won’t reset

Contacts table

Name

Email

Number

Role

Hadi Zaatiti

hz3752@nyu.edu

+971 56 275 4921

Research Scientist

Lawrence Torres

ljt7767@nyu.edu

NA

NA

Qiang Zhang

qz19@nyu.edu

NA

NA

QD Helium Recovery

heliumrecovery@qd-europe.com

NA

NA

QD Konstantin Voigt

voigt@qd-europe.com

NA

NA

QD Tobias Adler

adler@qd-europe.com

NA

NA

Ahmed Ansari

aa7703@nyu.edu

NA

Helium store manager (Primary contact for getting Helium tanks)

Mohammad Rakib

mr5527@nyu.edu

NA

Logistics and Sanitation Coordinator